Portaflow PT500 with Data Logger and Sensors

Portaflow PT500 with Data Logger and Sensors
Portaflow PT500 with Data Logger and Sensors
Price And Quantity
  • 1
  • Unit/Units
    Product Description

    Portaflow PT500 with Data Logger and Sensors for a wide range of Pipe Sizes

     

    ·      With two pairs of Transducers, the PT500 works on a wide range of pipe sizes from 0.51 to 78” (13 to 2000 mm) pipe diameters.

    ·      It includes a 98,000point data logger with USB and RS232 outputs.

    ·      The simple, quick-start calibration system calculates sensor separation distance based on pipe diameter and pipe wall material.

    ·      Flow is displayed in your choice of engineering units.

    ·      Data can be stored in selectable intervals from 5 seconds to 1 hour and displayed on the instrument in text or graph format.

    ·      The PT500 includes a watertight IP67 rated carrying case with foam padded compartments for the flowmeter components.

    ·      Compact, rugged and reliable, the PT500 is designed for accurate performance in a wide range of industrial applications and environments.

     

    Fluid/Application

    Recommended for clean liquids with less than 2% solids or gas bubbles

    Flow Rate Range

    0.65 to 40 ft/sec (0.2 to 12 m/sec)

    Pipe Size

    ‘A’ Transducers - 0.51" to 4.5" Outside Diameter (13 mm to 115 mm)

    ‘B’ Transducers - 2” to 78” Outside Diameter (50 mm to 2000 mm)

    Pipe Wall Thickness

    0.039 - 2.95” (1 - 75 mm)

    Pipe Materials

    Any sonic conducting material including carbon steel, stainless steel, PVC, PVDF, fiberglass, galvanized steel, mild steel, glass, copper, brass and pipes with bonded liners including epoxy, rubber and Teflon

    Display

    Backlit 64 x 240 pixel LCD. Displays flow rates in m3, liters, or gallons (per minute or hour), and flow velocity in feet per second or meters per second.

    Displays positive and negative flow and data log files in graphic or text modes.

    Totalizer

    12 digits - displays separate forward and reverse flow totals

    Power Input

    Built-in rechargeable NiMH battery powers the Portaflow for 16 hours continuously. The external charger operates from 90-264 VAC power input.

    Outputs

    0/4-20mA, 5V pulse (1 pulse/sec maximum)

    RS232 and USB for data transfer to a Windows PC

    Data Logger

    programmable, 98,000 data point capacity, time and date stamped values.

    Download to a PC or display onscreen

    Electronics Operating Temperature

    32° to 122°F (0° to 50°C). Storage Temperature: 14° to 140°F (-10° to 60°C)

    Carry Case

    rated IP67 with protective moulded foam insert

    Accuracy

    ±2% of reading in most applications, Repeatability: ±0.5%

    Calibration

    Built-in 16-key programming with user-friendly calibration menu. Stores calibration parameters for up to 20 different sites.

    Approvals

    CE (conforms to BS EN 61010, BS EN 61326 - 1:2006, BS EN 61326-2-3:2006). Charger is CE and UL approved. The Portaflow PT500 is certified for use in hazardous rated locations.

    Contact Us

    D - 998, First Floor, Palam Extn., Sector-7, New Delhi, Delhi, 110077, India
    Phone :+918068090079